SM8050
Magnetic-inductive flow meter
SMR11GGX10KG/US
ifm efector, inc. • 1100 Atwater Drive • Malvern • PA 19355 — We reserve the right to make technical alterations without prior notice. — EN-US — SM8050-02 — 30.06.2016 —
Outputs
Total number of outputs
1
Output signal
analog signal; IO-Link; (configurable)
Permanent current rating of
switching output DC
[mA]
250
Number of analog outputs
1
Analog current output
[mA]
4...20
Max. load
[Ω]
500
Overload protection
yes
Measuring/setting range
Measuring range
[l/min]
0.2...100
Accuracy / deviations
Flow monitoring
Accuracy (in the measuring
range)
± (0,8 % MW + 0,5 % MEW)
Repeatability
± 0,2% MEW
Reaction times
Flow monitoring
Response time
[s]
0.15; (dAP = 0, T19)
Interfaces
Communication interface
IO-Link
Transmission type
COM2 (38,4 kBaud)
IO-Link revision
1.1
SDCI standard
IEC 61131-9
Profiles
Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis
SIO mode
yes
Required master port class
A
Process data analogue
2
Min. process cycle time
[ms]
3
Supported DeviceIDs
Type of operation
DeviceID
default
577
Operating conditions
Ambient temperature
[°C]
-10...60
Storage temperature
[°C]
-25...80
Protection
IP 67
Tests / approvals
EMC
DIN EN 60947-5-9
Shock resistance
DIN IEC 68-2-27
20 g (11 ms)
Vibration resistance
DIN IEC 68-2-6
5 g (10...2000 Hz)
MTTF
[years]
167
Pressure equipment directive
sound engineering practice; can be used for group 2 fluids; group 1 fluids on request



Html Pages

1 2 3 4


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com