Electronic Components Datasheet Search
  English  ▼

X  

CPS76D Datasheet(PDF) 2 Page - Endress+Hauser Group Services AG

Part # CPS76D
Description  Memosens CPS76D
PDF  12 Pages
Scroll/Zoom Zoom In 100%  Zoom Out
Manufacturer  EH [Endress+Hauser Group Services AG]
Direct Link  https://www.endress.com/
Logo EH - Endress+Hauser Group Services AG

CPS76D Datasheet(HTML) 2 Page - Endress+Hauser Group Services AG

  CPS76D Datasheet HTML 1Page - Endress+Hauser Group Services AG CPS76D Datasheet HTML 2Page - Endress+Hauser Group Services AG CPS76D Datasheet HTML 3Page - Endress+Hauser Group Services AG CPS76D Datasheet HTML 4Page - Endress+Hauser Group Services AG CPS76D Datasheet HTML 5Page - Endress+Hauser Group Services AG CPS76D Datasheet HTML 6Page - Endress+Hauser Group Services AG CPS76D Datasheet HTML 7Page - Endress+Hauser Group Services AG CPS76D Datasheet HTML 8Page - Endress+Hauser Group Services AG CPS76D Datasheet HTML 9Page - Endress+Hauser Group Services AG Next Button
Zoom Inzoom in Zoom Outzoom out
 2 / 12 page
background image
Memosens CPS76D
2
Endress+Hauser
Function and system design
Measuring principle
pH measurement
The pH value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. The
membrane glass of the electrode supplies an electrochemical potential which is dependent upon the
pH value of the medium. This potential is generated by the selective penetration of H+ ions through
the outer layer of the membrane. An electrochemical boundary layer with an electric potential forms
at this point. An integrated Ag/AgCl reference system serves as the required reference electrode.
The transmitter converts the measured voltage into the corresponding pH value using the Nernst
equation.
ORP measurement
The ORP potential is a unit of measurement for the state of equilibria between oxidizing and
reducing components of a medium. The ORP is measured using a platinum or gold electrode instead
of the pH-sensitive glass membrane. Analog to the pH measurement, an integrated Ag/AgCl
reference system is used as a reference electrode.
rH measurement
The rH value is defined as the negative logarithm of partial pressure of hydrogen in a solution. The
pH value and ORP value of a solution must be measured simultaneously to calculate the rH value.
The value is calculated using the following equation:
rH = 2 . (mV/S) + 2 pH
pH = measured pH value
mV = measured ORP value in mV + 207 mV (Ag/AgCl system)
S = slope of pH electrode
The rH value is an indicator of the oxidation or reducing ability of a process solution. The measuring
range is from 0 to 42. A process medium with rH values between 0 and 9 has a high reducing ability;
a low reducing ability with rH 9-17; an indeterminate ability with rH 17-25; a low oxidation ability
with rH 25-34, and a high oxidation ability with rH 34-42.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com