  6027 Datasheet HTML 1Page - Christian Bürkert GmbH & Co. KG 6027 Datasheet HTML 2Page - Christian Bürkert GmbH & Co. KG 6027 Datasheet HTML 3Page - Christian Bürkert GmbH & Co. KG 6027 Datasheet HTML 4Page - Christian Bürkert GmbH & Co. KG 6027 Datasheet HTML 5Page - Christian Bürkert GmbH & Co. KG 6027 Datasheet HTML 6Page - Christian Bürkert GmbH & Co. KG 6027 Datasheet HTML 7Page - Christian Bürkert GmbH & Co. KG 6027 Datasheet HTML 8Page - Christian Bürkert GmbH & Co. KG 6027 Datasheet HTML 9Page - Christian Bürkert GmbH & Co. KG Next Button
Zoom Inzoom in Zoom Outzoom out
 5 / 26 page
background image
Type 8745
5 | 26
Visit product website 
Performance data
Nominal flow range (Q
N)
MFC: 20…1500 l/min (N
2)
MFMs ≤ 2500 l/min (N
2)
Maximum operating pressure
(overpressure to atmospheric pressure)
10 bar, optional up to 25 bar (MFM)
For MFCs, the maximum operating pressure depends on the medium and nominal valve
size
Measuring accuracy 2.)
± 1.5 % of reading ± 0.3 % FS (under calibration conditions)
Repeatability
± 0.1 % FS
Measuring span
1:50
Settling time (MFC)/Response time (MFM) (t
95 %)
< 500 ms
Electrical data
Power consumption 3.)
Max. 4 W (as MFM)
Max. 12.5…31.5 W (as MFC, depending on proportional valve type)
Approvals and conformities
Protection class
IP20
North America (USA/Canada)
Further information can be found in chapter “2.4. North America (USA/Canada)” on
page 6.
Foods and Beverages/Hygiene
Further information can be found in chapter “2.5. Foods and beverages/Hygiene” on
page 6.
Oxygen
Further information can be found in chapter “2.6. Oxygen” on page 6.
1.) For a detailed description of LED colours see operating instructions Type 8745 
2.) If the operating medium is different from the calibration medium, the actual measurement accuracy might vary from the indicated value. If the operating medium is natural gas, the
measurement accuracy depends on the composition of the natural gas, which can vary depending on the origin and season.
3.) Information in relation to the typical power consumption (at + 23 °C ambient temperature, nominal flow and 30 min closed-loop control mode). Information according to UL 61010 - 1
may vary (see operating instructions Type 8745 ).
1.3. Variant with electromotive proportional valve
Type 8745 with electromotive proportional valve is especially suitable for applications with high inlet pressures of up to 22 bar or high flow
rates (at a low pressure drop). The motor’s power consumption to hold a specific opening position is nearly zero. This key feature can
reduce the energy consumption of a plant dramatically. Without electrical power the valve remains in its current position.
The maximum duty cycle of the motor depends on the ambient temperature. The duty cycle does not refer to the duty cycle of the device
but to the duty cycle of the motor. The motor is not switched on unless the valve is to move. Frequent set-point value input changes will
drastically increase the motor’s duty cycle. Further information about the derating curve can be found in the chapter “6.3. Derating
diagram for electromotive variants” on page 23.
Product properties
Dimensions
Further information can be found in chapter “1.3. Variant with electromotive propor-
tional valve” on page 5.
Total weight
Approx. 1670 g (aluminium, valve Type 3280)
Approx. 2940 g (stainless steel 1.4404/316L, valve Type 3280)
LED display 1.)
With MFM: RGB LED according to NAMUR NE107
With valve: RGB LED to indicate valve opening
Control valve (proportional valve)
Normally closed
Valve orifice
2…20 mm
K
VS value range
0.5…7.8 m3/h
Performance data
Nominal flow range (Q
N)
20…2500 l/min (N
2)
Maximum operating pressure
(overpressure to atmospheric pressure)
22 bar (MFM)
For MFCs, the maximum operating pressure depends on the medium and nominal valve
size
Measuring accuracy
± 2 % of reading ± 0.5 % FS (under calibration conditions
Repeatability
± 0.5 % FS
Measuring span
1:50
Settling time (MFC)/Response time (MFM)
(t
95 %)
2.)
< 5 s
Electrical data
Power consumption
Max. 4 W (as MFM)
Max. 12 W (as MFC) 3.)
Approvals and conformities
Degree of protection
IP20
North America (USA/Canada)
Further information can be found in chapter “2.4. North America (USA/Canada)” on
page 6.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com