Electronic Components Datasheet Search
  English  ▼

X  

PMP105 Datasheet(PDF) 2 Page - TT Electronics.

Part # PMP105
Description  PMP105 Medical Power Supplies (90-105W)
PDF  3 Pages
Scroll/Zoom Zoom In 100%  Zoom Out
Manufacturer  TTELEC [TT Electronics.]
Direct Link  http://www.ttelectronics.com/
Logo TTELEC - TT Electronics.

PMP105 Datasheet(HTML) 2 Page - TT Electronics.

  PMP105 Datasheet HTML 1Page - TT Electronics. PMP105 Datasheet HTML 2Page - TT Electronics. PMP105 Datasheet HTML 3Page - TT Electronics.  
Zoom Inzoom in Zoom Outzoom out
 2 / 3 page
background image
© TT Electronics plc
General Note
All data sheets are subject to change without notice.
ISS.2 14/07/2020 Page: 2
TT Electronics | Power Partners, Inc
43 Broad Street Suite B206, Hudson, MA 01749, USA. t: +1 (978) 567-9600
Specifications
Safety Standards & EMC Specifications
Safety Standard Approvals
UL ES 60601-1, CSA C22.2 No. 60601-1 File No. E178020
TÜV EN 60601-1
EMI Standard
EN55011, FCC & VCCI Class B conducted and radiated
EMC Performance
EN61000-3-2: Harmonic distortion, class A and D
EN61000-3-3: Line flicker
EN61000-4-2: ESD, ±15 KV air and ±8 KV contact
EN61000-4-3: Radiated immunity, 10 V/m
EN61000-4-4: Fast transient/burst, ±2 KV
EN61000-4-5: Surge, ±1 KV diff., ±2 KV com
EN61000-4-6: Conducted immunity, 10 Vrms
EN61000-4-8: Magnetic field immunity, 30 A/m
EN61000-4-11: Voltage dip immunity, 30% reduction for 500 ms,
100% reduction for 10 ms
*Consult with TT Electronics for information on additional country safety approvals
Input Specifications
Input Voltage Range
90-264 VAC
Input Frequency Range
47 to 63Hz
Input Current
1.4 A (rms) for 115 VAC, 0.7 A (rms) for 230 VAC
Earth Leakage Current
200 μA max. @ 264 VAC, 63Hz
Touch Current
100 μA max. @ 264 VAC, 63Hz
Power Factor
>0.9
Output Specifications
Ripple and Noise
1% peak to peak maximum at the full load
Overvoltage Protection
Provided and set at 115-140% of its nominal output voltage
Overcurrent Protection
Protected to short circuit conditions
Temperature Coefficient
±0.04% /°C maximum
Transient Response
Maximum excursion of 4% or better on all models, recovering to 1%
of final value within 500 us after a 25% step load change
Environmental Specifications
Operating Temperature
0°C to +60°C (See Derating)
Storage Temperature
-40°C to +85°C
Relative Humidity
5% to 95% non-condensing
Temperature derating
Derate from 100% at +40°C linearly to 50% at +60°C
General Specifications
Switching Frequency
70-140KHz
Power Factor
0.98 Typical at 115 VAC
Efficiency
87% min. at full load
Hold-up Time
10ms minimum at 110 VAC
Line Regulation
±0.5% maximum at full load
Inrush Current
80 A @ 115 VAC or 120 A @ 230 VAC, at 25℃ cold start
Withstand Voltage
5600 VDC from input to output (2 MOPP)
2100 VDC from input to ground (1 MOPP)
700 VDC from output to ground
(To verify AC strength, get correct test method to avoid power sup-
ply damage.)
For Class II models, 4000 VAC from input to output
MTBF
150,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK
-217F



Html Pages

1 2 3


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp