Electronic Components Datasheet Search
  English  ▼

X  

ADS1259IPW Datasheet(PDF) 24 Page - Texas Instruments

Part # ADS1259IPW
Description  Industrial, 14kSPS, 24-Bit Analog-to-Digital Converter with Low-Drift Reference
PDF  51 Pages
Scroll/Zoom Zoom In 100%  Zoom Out
Manufacturer  TI2 [Texas Instruments]
Direct Link  https://www.ti.com
Logo TI2 - Texas Instruments

ADS1259IPW Datasheet(HTML) 24 Page - Texas Instruments

Back Button ADS1259IPW Datasheet HTML 20Page - Texas Instruments ADS1259IPW Datasheet HTML 21Page - Texas Instruments ADS1259IPW Datasheet HTML 22Page - Texas Instruments ADS1259IPW Datasheet HTML 23Page - Texas Instruments ADS1259IPW Datasheet HTML 24Page - Texas Instruments ADS1259IPW Datasheet HTML 25Page - Texas Instruments ADS1259IPW Datasheet HTML 26Page - Texas Instruments ADS1259IPW Datasheet HTML 27Page - Texas Instruments ADS1259IPW Datasheet HTML 28Page - Texas Instruments Next Button
Zoom Inzoom in Zoom Outzoom out
 24 / 51 page
background image
STARTPin
DRDY
Halted
Single
Conversion
Halted
Single
Conversion
OR
OR
START
START
DRDY
DOUT
Settled
Data
START
Pin
START
Command
or
t
SET
(1)
7thFallingSCLKEdgeofOpcode
V
=AINP
AINN
-
IN
SettledV
IN
ADS1259
SBAS424D
– JUNE 2009 – REVISED AUGUST 2011
www.ti.com
Pulse Control Mode (PULSE Bit = 1)
Settling Time Using START
In the Pulse Control mode, the ADS1259 performs a
When START goes high (via pin or command) a
single conversion when either the START pin is taken
delay may be programmed before the conversion
high or when the START command is sent. As seen
filter cycle begins. The programmable delay may be
in Figure 52, DRDY goes high when the conversion is
useful to provide time for external circuits (such as
started. When the conversion is complete, DRDY
after an external signal mux change), before the
goes low and further conversions are halted. To start
reading is started. Register bits DELAY[2:0] set the
a new conversion, transition the START pin back to
initial delay time as shown in Table 8.
high, or transmit the START opcode again.
Table 8. Initial START Delay
DELAY[2:0]
tDELAY (tCLK)
tDELAY (µs)
(1)
000
0
0
001
64
8.68
010
128
17.4
011
256
34.7
100
512
69.4
101
1024
139
110
2048
278
111
4096
556
(1) START opcode command takes effect on the seventh SCLK
falling edge.
(1) fCLK = 7.3728MHz.
Figure 52. Pulse Control Mode
After the programmable delay, the digital filter is reset
and a new conversion is started. DRDY goes low
when data are ready. There is no need to ignore or
CONVERSION SETTLING TIME
discard data; the data are completely settled. The
total time to perform the first conversion is the sum of
The ADS1259 features a digital filter architecture in
the programmable delay time and the settling of the
which settling time can be traded for wide filter
digital filter. That is, the value of Table 8 and Table 9
notches, resulting in improved line-cycle rejection.
combined. Figure 53 shows the timing and Table 9
This trade-off is determined by the selection of the
shows the settling time with programmable delay
sinc1 or sinc2 filter. The sinc1 filter settles in a single
equal to '0'.
cycle while the sinc2 filter provides wide-width filter
notches. The settling time of the ADS1259 is different
Table 9. Settling Time Using START
if START is used to begin conversions or if the
ADS1259 is free-running the conversions. These
SETTLING TIME (tSET) (ms)
(1)
DATA RATE
modes are explained in the Settling Time Using
(SPS)
sinc1
sinc2
START
and
Settling
Time
While
Continuously
10
100
200
Converting sections.
16.6
60.3
120
50
20.3
40.4
60
17.0
33.7
400
2.85
5.42
1200
1.18
2.10
3600
0.632
0.980
14,400
0.424
0.563
(1) fCLK = 7.3728MHz, DELAY[2:0] = 000.
(1) tSET = initial start delay plus the new conversion cycle time.
Figure 53. Data Retrieval Time After START
24
Copyright
© 2009–2011, Texas Instruments Incorporated



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com