  N8973A Datasheet HTML 1Page - Agilent(Hewlett-Packard) N8973A Datasheet HTML 2Page - Agilent(Hewlett-Packard) N8973A Datasheet HTML 3Page - Agilent(Hewlett-Packard) N8973A Datasheet HTML 4Page - Agilent(Hewlett-Packard) N8973A Datasheet HTML 5Page - Agilent(Hewlett-Packard) N8973A Datasheet HTML 6Page - Agilent(Hewlett-Packard) N8973A Datasheet HTML 7Page - Agilent(Hewlett-Packard) N8973A Datasheet HTML 8Page - Agilent(Hewlett-Packard) N8973A Datasheet HTML 9Page - Agilent(Hewlett-Packard) Next Button
Zoom Inzoom in Zoom Outzoom out
 2 / 12 page
background image
1. The N8974A and N8975A models have a mechanical switch fitted. This switch allows the analyzers to change between the 10 MHz to 3.0 GHz
and the 3.0 GHz to 6.7 GHz frequency range on the N8974A. On the N8975A, the switch allows the change between the 10 MHz to 3.0 GHz and the 3.0 GHz to
26.5 GHz frequency range. If the current measurement frequency range crosses the 3.0 GHz point, the switch will operate. The mechanical switch has a limited
number of cycles over which it is reliable. To maximize the reliable life of the switch, switching over the 3.0 GHz frequency point should be kept to a minimum.
2. Temperature stability on the standard frequency reference is achieved 60 minutes after the analyzer is turned on.
3. Option 1D5 recommended for applications requiring high frequency stability.
4. Parts per million (10-6).
5. Tuning accuracy is dependent on measurement bandwidth.
Specifications
Specifications are only valid for the stated operating frequency, and apply over
0° C to +55° C unless otherwise noted. The analyzer will meet its specifications
after 2 hours of storage within the operating temperature range, 60 minutes after
the analyzer is turned on, with Alignment running. A user calibration is required
before corrected measurements can be made.
Frequency
Frequency range1
N8973A
10 MHz to 3.0 GHz
N8974A
10 MHz to 6.7 GHz
N8975A
10 MHz to 26.5 GHz
Measurement bandwidth (nominal)
N8973A, N8974A, N8975A
4 MHz, 2 MHz, 1 MHz, 400 kHz, 200 kHz, 100 kHz
Frequency reference
Standard2
Opt.1D53
Aging
± < 2 ppm4/year
± < 0.1 ppm/year
Temperature stability
± < 6 ppm
± < 0.01 ppm
Settability
± < 0.5 ppm
± < 0.01 ppm
Tuning accuracy 5 (start, stop, center, marker)
4 MHz measurement bandwidth (default on all models of noise figure analyzer)
Frequency
Error
10 MHz – 3.0 GHz
± < reference error + 100 kHz
3.0 GHz – 26.5 GHz
± < reference error + 400 kHz
< 4 MHz- measurement bandwidth (functionality not present in N8972A)
Frequency
Error
10 MHz – 3.0 GHz
< ± reference error + 20 kHz
3.0 GHz – 26.5 GHz
< ± reference error + 20% of measurement bandwidth
2



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com