Electronic Components Datasheet Search
  English  ▼

X  

DPO3FLEX Datasheet(PDF) 13 Page - List of Unclassifed Manufacturers

Part # DPO3FLEX
Description  Mixed Signal Oscilloscopes
PDF  22 Pages
Scroll/Zoom Zoom In 100%  Zoom Out
Manufacturer  ETC2 [List of Unclassifed Manufacturers]
Direct Link  
Logo ETC2 - List of Unclassifed Manufacturers

DPO3FLEX Datasheet(HTML) 13 Page - List of Unclassifed Manufacturers

Back Button DPO3FLEX Datasheet HTML 9Page - List of Unclassifed Manufacturers DPO3FLEX Datasheet HTML 10Page - List of Unclassifed Manufacturers DPO3FLEX Datasheet HTML 11Page - List of Unclassifed Manufacturers DPO3FLEX Datasheet HTML 12Page - List of Unclassifed Manufacturers DPO3FLEX Datasheet HTML 13Page - List of Unclassifed Manufacturers DPO3FLEX Datasheet HTML 14Page - List of Unclassifed Manufacturers DPO3FLEX Datasheet HTML 15Page - List of Unclassifed Manufacturers DPO3FLEX Datasheet HTML 16Page - List of Unclassifed Manufacturers DPO3FLEX Datasheet HTML 17Page - List of Unclassifed Manufacturers Next Button
Zoom Inzoom in Zoom Outzoom out
 13 / 22 page
background image
MIL-STD-1553 (optional)
Trigger on Sync, Word Type 1 (Command, Status, Data), Command Word (set RT Address, T/R, Sub-address/Mode, Data Word
Count/Mode Code, and Parity individually), Status Word (set RT Address, Message Error, Instrumentation, Service Request Bit,
Broadcast Command Received, Busy, Subsystem Flag, Dynamic Bus Control Acceptance (DBCA), Terminal Flag, and Parity
individually), Data Word (user-specified 16-bit data value), Error (Sync, Parity, Manchester, Non-contiguous data), Idle Time
(minimum time selectable from 2 µs to 100 µs; maximum time selectable from 2 µs to 100 µs; trigger on < minimum, > maximum,
inside range, outside range). RT Address can be further specified to trigger on =, ≠, <, >, ≤, ≥ a particular value, or inside or
outside of a range.
I2S/LJ/RJ/TDM (optional)
Trigger on Word Select, Frame Sync, or Data. Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a specific data value, or
inside or outside of a range. Maximum data rate for I2S/LJ/RJ is 12.5 Mb/s. Maximum data rate for TDM is 25 Mb/s.
Parallel (available on MSO
models only)
Trigger on a parallel bus data value. Parallel bus can be from 1 to 16 bits (from the digital channels) plus 2 or 4 bits (from the
analog channels) in size. Binary and Hex radices are supported.
Acquisition system
Acquisition Modes
Sample
Acquire sampled values.
Peak Detect
Captures glitches as narrow as 2 ns at all sweep speeds.
Averaging
From 2 to 512 waveforms included in average.
Envelope
Min-max envelope reflecting Peak Detect data over multiple acquisitions.
Hi Res
Real-time boxcar averaging reduces random noise and increases vertical resolution.
Roll
Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div.
Waveform measurements
Cursors
Waveform and Screen.
Automatic measurements
29, of which up to four can be displayed on-screen at any one time. Measurements include: Period, Frequency, Delay, Rise Time,
Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase, Positive
Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle RMS, Positive
Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Area and Cycle Area.
Measurement statistics
Mean, Min, Max, Standard Deviation.
Reference levels
User-definable reference levels for automatic measurements can be specified in either percent or units.
Gating
Isolate the specific occurrence within an acquisition to take measurements on, using either the screen, or waveform cursors.
Power measurements (optional)
Power Quality Measurements
VRMS, VCrest Factor, Frequency, IRMS, ICrest Factor, True Power, Apparent Power, Reactive Power, Power Factor, Phase Angle.
Switching loss measurements
Power loss
Ton, Toff, Conduction, Total.
Energy loss
Ton, Toff, Conduction, Total.
Harmonics
THD-F, THD-R, RMS measurements. Graphical and table displays of harmonics. Test to IEC61000-3-2 Class A and MIL-
STD-1399, Section 300A.
Ripple measurements
VRipple and IRipple.
Modulation Analysis
Graphical display of +Pulse Width, –Pulse Width, Period, Frequency, +Duty Cycle, and –Duty Cycle modulation types.
1
Trigger selection of Command Word will trigger on Command and ambiguous Command/Status words. Trigger selection of Status Word will trigger on Status and ambiguous Command/Status words.
MSO3000 Series and DPO3000 Series Mixed Signal Oscilloscopes
Trigger system
www.tektronix.com 13



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com