Back Button U2902A Datasheet HTML 17Page - Keysight Technologies U2902A Datasheet HTML 18Page - Keysight Technologies U2902A Datasheet HTML 19Page - Keysight Technologies U2902A Datasheet HTML 20Page - Keysight Technologies U2902A Datasheet HTML 21Page - Keysight Technologies U2902A Datasheet HTML 22Page - Keysight Technologies U2902A Datasheet HTML 23Page - Keysight Technologies U2902A Datasheet HTML 24Page - Keysight Technologies U2902A Datasheet HTML 25Page - Keysight Technologies  
Zoom Inzoom in Zoom Outzoom out
 21 / 25 page
background image
21
System accuracy specifications
The U2802A system accuracy specifications are shown in Table 1, Table 2, and Table 3. These measurements are derived from
the U2802A and DAQ input accuracy specifications without including the thermocouple error. Refer to “Calculating System
Accuracy” section in the Keysight U2802A 31-Channel Thermocouple Input Device User’s Guide for calculation methodology.
Table 1. Measurement accuracy of the U2355A and U2356A at 23 °C ± 5 °C
Thermocouple measurement accuracy (U2355A, U2356A at 23 °C ± 5 °C)
T/C type
ITS-90 Temperature range (°C)
Optimum measurement
range(°C)
Without
averaging
(± °C)
50 points
averaging
(± °C)
500 points
averaging
(± °C)
Low
High
Low
High
B
0
1820
1100
1820
1.9
1.2
1.0
400
1100
4.4
2.5
2.0
E
–270
1000
–150
1000
1.7
1.6
1.6
–200
–150
2.4
2.3
2.3
J
–210
1200
–150
1200
1.6
1.5
1.5
–210
–150
2.7
2.6
2.5
K
–270
1372
–100
1200
1.5
1.4
1.4
–200
–100
2.7
2.6
2.6
N
–270
1300
–100
1300
1.5
1.3
1.3
–200
–100
3.0
2.7
2.6
R
–50
1768
300
1760
2.0
1.4
1.3
–50
300
5.0
3.1
2.6
S
–50
1768
400
1760
2.1
1.6
1.4
–50
400
4.5
2.8
2.4
T
–270
400
–100
400
1.5
1.4
1.4
–200
–100
2.7
2.5
2.5
Table 2. Measurement accuracy of the U2355A at 0 to 18 °C and 28 to 45 °C
Thermocouple measurement accuracy (U2355A at 0 to 18 °C and 28 to 45 °C)
T/C type
ITS-90 Temperature range (°C)
Optimum measurement range
(°C)
Without
|averaging
(± °C)
50 points
averaging
(± °C)
500 points
averaging
(± °C)
Low
High
Low
High
B
0
1820
1100
1820
3.4
2.4
2.2
400
1100
7.5
3.6
2.2
E
–270
1000
–150
1000
2.7
2.6
2.5
–200
–150
3.8
3.6
3.6
J
–210
1200
–150
1200
2.5
2.4
2.4
–210
–150
4.2
4.0
3.9
K
–270
1372
–100
1200
2.9
2.8
2.8
–200
–100
4.3
4.0
3.9
N
–270
1300
–100
1300
2.6
2.5
2.5
–200
–100
4.9
4.2
4.0
R
–50
1768
300
1760
3.8
3.1
3.0
–50
300
8.5
4.6
3.3
S
–50
1768
400
1760
4.2
3.4
3.2
–50
400
7.7
4.2
3.1
T
–270
400
–100
400
2.4
2.2
2.2
–200
–100
4.3
4.3
3.9



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com