33 / 42 page ![]() 2 Service program features 5140241-100 rev 3 Instrumentarium Dental 27 2.15 Sr 88 CAL: Set cephalostat alignment value (film units only) To align the THA with the cephalostat. The adjustment value set here affects the rotational angle of the THA, and thereby the horizontal position of the x-ray beam on the film. A positive adjustment value moves the beam to the right and vice versa. The value is shown in centimeters. Usage: 1. Select Sr 88 CAL. The current adjustment value is shown on the time/dose display. 2. Using the left and right buttons, modify the adjustment value as appropriate. 3. Press OK to exit the program. 2.16 Sr 89 COP: Set country options Lets you customize the unit according to regional regulations and user expectations. Lists a number of options, each having a number (shown in the kV field) and a short name (shown in the mA field). The value associated with the option is shown in the time/dose field. The options available are described below. Approved: Äärynen Teemu 2008-05-19 06:07 Reviewed: Rintamäki Markus 2008-05-16 15:20 Approved D500293, 3 |
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |