Back Button OC200D Datasheet HTML 34Page - List of Unclassifed Manufacturers OC200D Datasheet HTML 35Page - List of Unclassifed Manufacturers OC200D Datasheet HTML 36Page - List of Unclassifed Manufacturers OC200D Datasheet HTML 37Page - List of Unclassifed Manufacturers OC200D Datasheet HTML 38Page - List of Unclassifed Manufacturers OC200D Datasheet HTML 39Page - List of Unclassifed Manufacturers OC200D Datasheet HTML 40Page - List of Unclassifed Manufacturers OC200D Datasheet HTML 41Page - List of Unclassifed Manufacturers OC200D Datasheet HTML 42Page - List of Unclassifed Manufacturers  
Zoom Inzoom in Zoom Outzoom out
 39 / 42 page
background image
2 Service program features
5140241-100 rev 3
Instrumentarium Dental
33
1) With manual collimator, collimator type checking is
disabled. With automatic collimator, this level disables
driving at the collimator and earholder position checking.
Usage:
1. Select Sr 92 Che.
2. Using the left and right button, select the desired level.
Select OFF to enable all checks and operations.
3. Press OK to exit the program. Checking is set to the
selected level.
2.20 Sr 93 BAL: Adjust beam alignment board
sensitivity
NOTE! See the OP200 installation manual.
For adjusting the sensitivity of the sensors on the beam
alignment board. It assists in driving the cephalostat
camera and secondary collimator into the correct position
and selecting appropriate exposure values.
This program has two modes of operation:
Test mode positions the unit according to the currently
selected channel.
Manual
mode
performs
an
exposure
without
movements.
The AEC density LEDs indicate the currently selected
channel. The midmost LED corresponds to the mid
channel on the Beam alignment board.
In units equipped with automatic collimator, the selected
collimator type (normal height / pediatric) is indicated by
the pediatric collimation LED.
Usage:
1. Select Sr 93 BAL.
2. Select the channel to be adjusted by pressing the up /
down buttons until one of the AEC density LEDs is
blinking. Change channel by pressing the left / right
buttons.
3. Select correct collimator.
Manual collimator: Move the cephalometric collimator
into position. Use copper filtration as appropriate for the
channel.
Approved: Äärynen Teemu 2008-05-19 06:07
Reviewed: Rintamäki Markus 2008-05-16 15:20
Approved
D500293, 3



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com