13 / 36 page ![]() MODEL R302/A . INSTRUCTIONS MANUAL MTR302/A -ING. - page E-1 E) COLLIMATOR CALIBRATION: WARNING: THE FOLLOWING PROCEDURES REQUIRE THAT X-RADIATION BE PRODUCED. TAKE ADEQUATE PRECAUTIONS TO SEE THAT NO PART OF THE HUMAN BODY IS EXPOSED TO X-RADIATION, DIRECT OR INDIRECT. Centering, X-ray beam Place a 35x43 cm. (14x17") cassette on the table top or other flat, horizontal surface and position the x-ray tube/collimator assembly with the focal spot at 1 meter (40") above and with the x-ray beam perpendicular to the cassette surface. Do not use equipment scales for reference, but measure the distance from focal spot to cassette surface. Note: if one meter FFD (SID) cannot be obtained, use the obtainable FFD (SID) that is closest to one meter and calculate the measurement tolerances as the appropriate percentages of the distance. If the x-ray beam cannot be oriented vertically, then make provision by using clamps, masking tape, etc. as required to place the test objects and image receptor at the specified FFD (SID) and perpendicular relative to the x-ray beam as described in the following procedures. • Use the collimator light to centre the cassette in the field. • Mark the location of the cassette with masking tape or other means so that it may be removed and replaced in the same position. • Place white paper on top of the cassette to provide maximum contrast for the light field. • Set the collimator to provide a field size of 35x35 cm. (14x14") at one meter (40"). • Activate the light field and use it to position 20 coins as shown in the diagram. • Position each pair of coins touching one another so that the inner coin is lighted as much as possible and the outer coin is lighted as little as possible. • The points of tangency will define the edges of the light field. The extra four coins in the lower left corner will provide the means for adjusting the film. • Set the technique factors at the x-ray generator to produce a density of about 1 (about 50 kVp, 5 mAs). • Make an exposure. Remove the cassette and process the film. Use the test film to check the alignments: described in the following paragraphs. |
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |