Back Button MAX86150 Datasheet HTML 35Page - Maxim Integrated Products MAX86150 Datasheet HTML 36Page - Maxim Integrated Products MAX86150 Datasheet HTML 37Page - Maxim Integrated Products MAX86150 Datasheet HTML 38Page - Maxim Integrated Products MAX86150 Datasheet HTML 39Page - Maxim Integrated Products MAX86150 Datasheet HTML 40Page - Maxim Integrated Products MAX86150 Datasheet HTML 41Page - Maxim Integrated Products MAX86150 Datasheet HTML 42Page - Maxim Integrated Products MAX86150 Datasheet HTML 43Page - Maxim Integrated Products Next Button
Zoom Inzoom in Zoom Outzoom out
 39 / 49 page
background image
Read Data Format
For the read operation, two I2C operations must be
performed. First, the slave ID byte is sent followed by the
I2C register that you wish to read. Then a REPEAT START
(Sr) condition is sent, followed by the read slave ID. The
MAX86150 then begins sending data beginning with the
register selected in the first operation. The read pointer
increments automatically, so the MAX86150 continues
sending data from additional registers in sequential order
until a STOP (P) condition is received. The exception to
this is the FIFO_DATA register, at which the read pointer no
longer increments when reading additional bytes. To read
the next register after FIFO_DATA, an I2C write command
is necessary to change the location of the read pointer.
Figure 7 and Figure 8 show the process of reading one
byte or multiple bytes of data respectively.
An initial write operation is required to send the read
register address.
Data is sent from registers in sequential order, starting
from the register selected in the initial I2C write operation.
If the FIFO_DATA register is read, the read pointer does
not automatically increment, and subsequent bytes of
data contain the contents of the FIFO.
Figure 7. Reading One Byte of Data from MAX86150
Figure 8. Reading Multiple Bytes of Data from the MAX86150
www.maximintegrated.com
Maxim Integrated │ 39
MAX86150
Integrated Photoplethysmogram and
Electrocardiogram Bio-Sensor Module For
Mobile Health
S
1
0
1
1
1
1
0
R/W
= 0
ACK
A7
A6
A5
A4
A3
A2
A1
A0
ACK
SLAVE ID
REGISTER ADDRESS
DATA BYTE
S = START CONDITION
Sr = REPEATED START CONDITION
P= STOP CONDITION
Sr
1
0
1
1
1
1
0
R/W
= 1
ACK
D7
D6
D5
D4
D3
D2
D1
D0
AM
DATA 1
DATA n-1
D7
D6
D5
D4
D3
D2
D1
D0
NACK
P
DATA n
D7
D6
D5
D4
D3
D2
D1
D0
AM
ACK = ACKNOWLEDGE BY THE RECEIVER
NACK = NOT ACKNOWLEDGE
AM = ACKNOWLEDGE BY THE MASTER



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com