Electronic Components Datasheet Search
  English  ▼

X  

CAL01 Datasheet(PDF) 4 Page - MultiDimension Technology Co.,Ltd.

Part # CAL01
Description  CAL01 – 800 GAUSS CALIBRATION MAGNET
PDF  5 Pages
Scroll/Zoom Zoom In 100%  Zoom Out
Manufacturer  MULTIDIMENSION [MultiDimension Technology Co.,Ltd.]
Direct Link  http://www.dowaytech.com/en/
Logo MULTIDIMENSION - MultiDimension Technology Co.,Ltd.

CAL01 Datasheet(HTML) 4 Page - MultiDimension Technology Co.,Ltd.

  CAL01 Datasheet HTML 1Page - MultiDimension Technology Co.,Ltd. CAL01 Datasheet HTML 2Page - MultiDimension Technology Co.,Ltd. CAL01 Datasheet HTML 3Page - MultiDimension Technology Co.,Ltd. CAL01 Datasheet HTML 4Page - MultiDimension Technology Co.,Ltd. CAL01 Datasheet HTML 5Page - MultiDimension Technology Co.,Ltd.  
Zoom Inzoom in Zoom Outzoom out
 4 / 5 page
background image
MultiDimension Technology Co., Ltd.
Page 4 of 5
Web: www.dowaytech.com/en, Email: info@dowaytech.com
MDT USB2705A
No.7 Guangdong Road, Zhangjiagang Free Trade Zone, Jiangsu, 215634, P.R. China
Ver. 1.0a
Axial Probe Calibration Procedure
1.
Press the “calibrate to field” button.
2.
Set the calibration field value to the magnetic field value indicated on the side of the calibration
assembly
3.
Make sure both sensitivity and offset checkboxes are selected
4.
Push the axial probe into the axial alignment hole to find the maximum positive (negative)
voltage reading, you can wiggle the probe some and move in and out of the calibration
assembly to verify the maximum value has been sampled.
5.
Hold the probe steady and pull the calibration assembly away.
6.
Flip the calibration magnet in order to reverse the magnetic field
7.
Push the axial probe into the axial alignment hole to find the maximum negative (positive)
voltage reading, you can wiggle the probe some and move in and out of the calibration
assembly to verify the maximum value has been sampled.
8.
Save the calibration in either the flash for future use or just write the calibration to the sensor
for immediate use only
9.
Exit the calibration window
10. Enter the digitizer window and turn on the low pass filter (LPF) if desired.
11. If the offset has not been adequately corrected, perform one of the offset calibration procedures
Motion of the Magnet and Probe
MultiDimension Technology Graphical Display



Html Pages

1 2 3 4 5


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com