Back Button 8757D Datasheet HTML 3Page - Agilent(Hewlett-Packard) 8757D Datasheet HTML 4Page - Agilent(Hewlett-Packard) 8757D Datasheet HTML 5Page - Agilent(Hewlett-Packard) 8757D Datasheet HTML 6Page - Agilent(Hewlett-Packard) 8757D Datasheet HTML 7Page - Agilent(Hewlett-Packard) 8757D Datasheet HTML 8Page - Agilent(Hewlett-Packard) 8757D Datasheet HTML 9Page - Agilent(Hewlett-Packard) 8757D Datasheet HTML 10Page - Agilent(Hewlett-Packard) 8757D Datasheet HTML 11Page - Agilent(Hewlett-Packard) Next Button
Zoom Inzoom in Zoom Outzoom out
 7 / 14 page
background image
7
Precision detector vs. power sensor absolute
power measurement accuracy
Accuracy terms differ depending on the test equip-
ment used to make absolute power measurements.
The following table simplifies and compares the
accuracy terms of an Agilent 8757 system (using
an 85037 series precision detector) and a power
meter (using a power sensor). A measurement
accuracy example is also provided.
Scalar terms
Equivalent power meter/sensor terms
Absolute power
Power reference uncertainty
accuracy at 50 MHz
Instrument linearity
Zero set
Noise
Frequency response
Sensor calibration factor uncertainty
Mismatch
Mismatch
Scalar analyzer vs. power meter
Absolute power measurement uncertainty examples
Assumptions:
• Measurement frequency = 10 GHz
• DUT input/output SWR = 1.5
• Power measurement range = +10 to –20 dBm
Uncertainty component
(see above table for
8757D Opt. 002/
EPM-4418B
equivalent power meter terms) 85037B
8485A
Absolute power accuracy
at 50 MHz (±dB)
0.11
0.09
Frequency response (±dB)
0.18
0.09
Mismatch (±dB)
0.18
0.12
Total (±dB)
0.47
0.30
Reflection measurement accuracy
Uncertainties due to calibration error and the
frequency response of the source, detectors, and
bridges are removed via open/short averaging.
The remaining uncertainties are primarily the sum
of directivity uncertainty, effective source match
uncertainty, and dynamic power accuracy. As shown
in the graphs below, directivity is the dominant error
term when measuring small reflected signals (high
return loss) and source match is dominant when
measuring large reflected signals (low return loss).
Effective source match vs. reflection uncertainty
Directivity vs. reflection uncertainty
Example calculation
The following example shows how to find the
uncertainty (excluding dynamic accuracy) in
measuring a 14-dB return loss (SWR = 1.5)
with an 85027A directional bridge at 10 GHz
(directivity = 40 dB, test port match = 1.25 SWR).
Uncertainty component
Uncertainty
Source match error
approximately ±0.2 dB
Directivity error
approximately ±0.4 dB
Total uncertainty
approximately ±0.6 dB



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com