  LG16 Datasheet HTML 1Page - List of Unclassifed Manufacturers LG16 Datasheet HTML 2Page - List of Unclassifed Manufacturers LG16 Datasheet HTML 3Page - List of Unclassifed Manufacturers LG16 Datasheet HTML 4Page - List of Unclassifed Manufacturers LG16 Datasheet HTML 5Page - List of Unclassifed Manufacturers LG16 Datasheet HTML 6Page - List of Unclassifed Manufacturers LG16 Datasheet HTML 7Page - List of Unclassifed Manufacturers LG16 Datasheet HTML 8Page - List of Unclassifed Manufacturers  
Zoom Inzoom in Zoom Outzoom out
 5 / 8 page
background image
www.sensirion.com
February 2015 V3
5/8
3 Communication with the Sensor
The OEM flow sensor LG16 shows bi-directional,
linear transfer characteristics. It can be ordered as
analog or digital output version. Please see the
following section for more details and the table on
page 7 for an overview of the available options.
Digital Sampling Time, 16 bit
74 ms
Digital Sampling Time, 9 bit
1 ms
3.1 Electrical Specifications
Table 4: DC Characteristics
Parameter
Min.
Typ.
Max.
Units
Power Supply DC, VDD
Digital version
4.0
5.0
12a
V
Analog version
6.0
7.0
12a
V
Operating Current,
VDD = 4.0-12 V, no load
6.8
mA
a
Use 9 V max. for best performance
3.2 Electrical Connector and Sensor Pinout
Connector Type: 5 pin Molex PCB Header 53261-
0590 (right angle). Assembled flat ribbon cable Molex
1.25 mm Pitch Receptacle Type 51021-0500
(PicoBlade™ 51021) included.
Pin
1
SDA (bi-directional)
2
SCL
3
VDD
4
GND
5
Analog out
3.3 Digital Communication via I²C-Bus
Digital communication between a master and the
LG16 sensor runs via the standard I²C-interface. The
physical interface consists of two bus lines, a data line
(SDA) and a clock line (SCL) which need to be
connected via pull-up resistors to the bus voltage of
the system.
These lines can be used on 3.3V or 5.0V level with a
clock frequency of 100 kHz. For the detailed
specifications of this I²C communication, please refer
to specific I²C Application Notes from Sensirion.
3.4 Analog Communication
The LG16 is available as analog version, with the
voltage output representing the current flow rate.
Analog Out Voltage Range
Negative full scale flow
0.2 V
Zero flow
2.5 V
Positive full scale flow
4.8 V
Minimum load resistance
10 k
The analog sensor’s output limit is 5V, thus the actual
flow rate output limit is lower than the corresponding
digital version’s output limit (noted in Table 1).
1 2
3 4
5



Html Pages

1 2 3 4 5 6 7 8


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com