4 / 8 page ![]() Condumax CLS12 4 Endress+Hauser 3. If using the sensor in the ultrapure water range: Work under air-evacuated conditions. This prevents CO2 in the air from dissolving in the water and increasing the conductivity by up to 3 µS/cm as a result of (weak) dissociation. Environment Ambient temperature range –20 to 60 °C (–4 to 140 °F) Storage temperature -25 to +80 °C (-10 to +180 °F) Degree of protection IP67 Process Process temperature -20 to 160 ˚C (-4 to 320 ˚F) Process pressure Without flow assembly CLA751 Up to 100 °C (212 °F): 1 to 41 bar (15 to 595 psi), absolute Up to 160 °C (320 °F): 1 to 7 bar (15 to 102 psi), absolute In flow assembly CLA751 1 to 13 bar (15 to 185 psi), absolute Temperature/pressure ratings 7 20 30 41 0 30 [°C] -4 32 212 102 290 435 - 0 2 [°F] 595 [psi] [bar] p (abs.) T 100 320 70 160 A0048118 4 Mechanical pressure-temperature resistance |
|
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|