  CM72 Datasheet HTML 1Page - Endress+Hauser Group Services AG CM72 Datasheet HTML 2Page - Endress+Hauser Group Services AG CM72 Datasheet HTML 3Page - Endress+Hauser Group Services AG CM72 Datasheet HTML 4Page - Endress+Hauser Group Services AG CM72 Datasheet HTML 5Page - Endress+Hauser Group Services AG CM72 Datasheet HTML 6Page - Endress+Hauser Group Services AG CM72 Datasheet HTML 7Page - Endress+Hauser Group Services AG CM72 Datasheet HTML 8Page - Endress+Hauser Group Services AG CM72 Datasheet HTML 9Page - Endress+Hauser Group Services AG Next Button
Zoom Inzoom in Zoom Outzoom out
 5 / 12 page
background image
Liquiline Compact CM72
Endress+Hauser
5
The lower voltage value in each case applies only to a load resistance of 0 Ohm.
NOTICE
The device does not have a power switch
‣ At the supply point, the power supply must be isolated from dangerous live cables by double or
reinforced insulation in the case of devices with a 24 V power supply.
Cable specification
Cable length:
• Max. 3 m (10 ft)
• Max. 7 m (23 ft)
• Max. 15 m (49 ft)
Overvoltage protection
IEC 61 000-4-4 and IEC 61 000-4-5 with +/- 1 kV
Performance characteristics
Resolution
Current output
< 5 µA
Repeatability
→ Documentation of the connected sensor
Response time
Current output
t90 = max. 500 ms for an increase from 0 to 20 mA
Tolerance
Current output
Typical measuring tolerances:
< ±20 µA (if current value = 4 mA)
< ±50 µA (for current values 4 to 20 mA)
at 25 ˚C (77˚ F) each
additional tolerance depending on the temperature:
< 1.5 µA/K
Environment
Ambient temperature range
–20 to 85 °C (–4 to 185 °F)
The maximum ambient temperature depends on the process temperature and the transmitter's
installation position.
‣ Make sure that the ambient temperature at the transmitter does not exceed 85 °C (185 °F).
Example for ambient conditions in Endress+Hauser assemblies:
• for open installation (without protective cover, i.e. free convection at the transmitter), e.g.
CPA442, CPA842
• for enclosed installation (with protective cover), e.g. CPA871, CPA875, CPA842
Tambient = max. 60 °C (140 °F)
Tprocess = max. 100 °C (212 °F), in continuous operation
Tprocess = max. 140 °C (284 °F), < 2h (for sterilization)



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com