Electronic Components Datasheet Search
  English  ▼

X  

BMM350 Datasheet(PDF) 14 Page - Bosch Sensortec GmbH

Part # BMM350
Description  3-axis magnetic sensor with high data rate
PDF  63 Pages
Scroll/Zoom Zoom In 100%  Zoom Out
Manufacturer  BOSCH [Bosch Sensortec GmbH]
Direct Link  https://www.bosch-sensortec.com
Logo BOSCH - Bosch Sensortec GmbH

BMM350 Datasheet(HTML) 14 Page - Bosch Sensortec GmbH

Back Button BMM350 Datasheet HTML 10Page - Bosch Sensortec GmbH BMM350 Datasheet HTML 11Page - Bosch Sensortec GmbH BMM350 Datasheet HTML 12Page - Bosch Sensortec GmbH BMM350 Datasheet HTML 13Page - Bosch Sensortec GmbH BMM350 Datasheet HTML 14Page - Bosch Sensortec GmbH BMM350 Datasheet HTML 15Page - Bosch Sensortec GmbH BMM350 Datasheet HTML 16Page - Bosch Sensortec GmbH BMM350 Datasheet HTML 17Page - Bosch Sensortec GmbH BMM350 Datasheet HTML 18Page - Bosch Sensortec GmbH Next Button
Zoom Inzoom in Zoom Outzoom out
 14 / 63 page
background image
Bosch Sensortec | BST-BMM350-DS001-27
14 | 63
© Bosch Sensortec GmbH 2023 | All rights reserved, also regarding any disposal, exploitation, reproduction, editing, distribution,
Document number: BST-BMM350-DS001-27
as well as in the event of applications for industrial property rights
5.1 Power modes, boot procedure
The BMM350 knows two major modes of operation, “normal mode” and “forced mode”. “suspend mode” brings the
device into minimal power consumption, where settings are retained and communication is possible while data
conversions are stopped.
After boot, the device will be in suspend mode, consuming the least current possible offered by this device.
Immediately after boot the API downloads the compensation coefficients from the BMM350 OTP (one-time
programmable memory).
This is handled automatically in the API function BMM350_init().
The boot phase must be terminated by writing 0x80 to OTP_CMD_REG (also done in BMM350_init).
From then on the OTP is inaccessible to the API unless a power reset or soft reset triggers another boot.
Power modes are selected using register PMU_CMD.pmu_cmd.
An API function “BMM350_set_powermode” is provided which handles power mode settings.
5.1.1 Power off
In Power off, VDD and/or VDDIO are unpowered/below brownout threshold and the device does not operate. When
only one of VDD or VDDIO is supplied, the magnetic sensor will still be in Power off. Power on/boot is performed
immediately after both VDD and VDDIO are raised above their reset release thresholds.
The sequence of raising VDD and VDDIO is left to the user, both voltages can be raised in any desired order.
If power is stable, the internal OTP becomes accessible (OTP access granted).
5.1.2 Suspend mode
Suspend mode is the default power mode of BMM350 after the chip is powered and has finished booting when using
BMM350_init API function. This function reads out the OTP which stores compensation coefficients.
Suspend mode can be reached on demand via sending a suspend mode request to register PMU_CMD.pmu_cmd, or
it is reached automatically after finishing a “forced mode” conversion.
Current consumption in suspend mode is minimal, so, this mode is useful for periods when data conversion is not
needed. Read and write of all registers is possible.
5.1.3 Normal mode
Normal mode is activated via register PMU_CMD.pmu_cmd.
An API function BMM350_set_powermode is provided which helps doing that.
In normal mode the internal oscillator triggers conversions with the chosen ODR (output data rate) with the accuracy
that can be expected from an RC oscillator.
The noise performance, tightly coupled with the current consumption, is controlled via the “averaging” setting, see
register PMU_CMD_AGGR_SET.avg . As the name indicates 1, 2, 4 or 8 data samples are generated internally and
averaged before the result is written into the data register. This improves the noise performance at the cost of current
consumption, which is increasing proportionally with the averaging factor setting.
Not all combinations of output data rate and averaging are possible, see Table 5.
An API function BMM350_set_performance is provided which helps setting the averaging/noise performance. The
function checks for the combination of ODR and averaging setting and chooses the highest possible averaging setting
if the desired averaging setting is too high.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com