  ES1C Datasheet HTML 1Page - Omron Electronics LLC ES1C Datasheet HTML 2Page - Omron Electronics LLC ES1C Datasheet HTML 3Page - Omron Electronics LLC ES1C Datasheet HTML 4Page - Omron Electronics LLC ES1C Datasheet HTML 5Page - Omron Electronics LLC ES1C Datasheet HTML 6Page - Omron Electronics LLC ES1C Datasheet HTML 7Page - Omron Electronics LLC ES1C Datasheet HTML 8Page - Omron Electronics LLC  
Zoom Inzoom in Zoom Outzoom out
 4 / 8 page
background image
ES1C
4
B Two-point Shift
1. Measuring the Temperature of the Measurement Object
Set the temperature of the measurement object to room
temperature and to the temperature during operation, and
check the values indicated by the connected device (A)
and the temperatures of the measurement object (B).
2. Shifting the Indicated Value
Use the following formula to calculate the upper limit and
lower limit of input scaling after the shift from the values Y1
and Y2 indicated by the connected device and tempera-
tures X1 and X2 of the measurement object.
(1) Scaling upper limit after shift (
°C)
(2) Scaling lower limit after shift (
°C)
Change the values to the scaling upper and lower limits
from the result considering the decimal point position. For
example, if the scaling upper limit after shift is 487.5 (
°C)
and the scaling lower limit after shift is
−12.5 (°C), the dec-
imal point position of the connected device will be set to the
first decimal position, and so the scaling upper limit will be
set to 4,875 and the scaling lower limit to
−125.
● Adjustment Example for the K3GN
A Shift Method
1. Measuring the Temperature of the Measurement Object
Use a thermometer (B) to measure the actual temperature
when using the measurement object (C).
2. Shifting the Display Value
Adjust the settings of the connected device after checking
the following value:
Temperature B (thermometer)
− Temperature A (connect-
ed device)
If temperature B minus temperature A is 10.0, adjust the
settings so that measurement value of the connected de-
vice is +10.0.
• Scaling display value 1 = 0 to 100
• Scaling display value 2 = 4,000 to 4,100
(The setting for the decimal point position is 0000.0, and so
the scaling set value will be increased by 100 for a dis-
played value of +10.0.)
B Two-point Shift
Use the teaching function of the K3GN to make adjustments
using the ES1C's actual analog input value and the actual tem-
perature. Set one of the two teaching points to room tempera-
ture and the other to the actual temperature of the
measurement object during operation.
1. Move the K3GN to the initial setting level.
2. Set the temperature of the object to be measured to room
temperature and set scaling input value 1 using teaching.
Next, set the temperature (B) of the thermometer to scal-
ing display value 1. (Point
A in the following figure. The
decimal point position for the K3GN is set to the first deci-
mal position, and so set 250 for 25.0
°C.)
3. Next, set the measurement object to the actual operating
temperature and set scaling input value 2 and scaling dis-
play value 2 in step 2. (A value of 1,000 is set to specify
100.0
°C for point B in the following graph.)
Value indicated by
connected device (A)
Temperature of mea-
surement object (B)
Room tem-
perature
Y1
X1
Temperature
during opera-
tion
Y2
X2
Before shift
After shift
Value indicated by connected device (A)
X1 room
temperature
(e.g., 25
°C)
X2 temperature
during operation
(e.g., 100
°C)
Value indicated after shift
(e.g., 100.0)
Value Y2 indicated before
shift (e.g., 90.0)
Value Y1 indicated before
shift (e.g., 30.0)
Value indicated after shift
(e.g., 25.0)
Temperature
displayed by
thermometer (B)
(
°C)
(
°C)
X2 X1
–
Y2 Y1
–
---------------------400 Y1
–
() X1
+
=
X2 X1
–
Y2 Y1
–
---------------------0 Y1
–
() X1
+
=
Before shift
After shift
Value indicated by connected device (A)
Measurement
object temperature
(e.g., 100
°C)
Value indicated after shift
(e.g., 100.0)
Value indicated before
shift (e.g., 90.0)
Temperature
displayed by
thermometer (B)
(
°C)
(
°C)
+10.0
0
A
B
Value indicated by connected device (A)
Measurement object at operating
temperature (e.g., 100
°C)
Measurement object at room
temperature (e.g., 25
°C)
Scaling input value 2
(e.g., 8.02 mA)
Scaling input value 1
(e.g., 5.01 mA)
Scaling display value 2
(e.g., 100.0
°C)
(
°C)
Scaling display value 1
(e.g., 25.0
°C)
Input value of
connected
device
(mA)



Html Pages

1 2 3 4 5 6 7 8


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com