Back Button ADPD7000 Datasheet HTML 15Page - Analog Devices ADPD7000 Datasheet HTML 16Page - Analog Devices ADPD7000 Datasheet HTML 17Page - Analog Devices ADPD7000 Datasheet HTML 18Page - Analog Devices ADPD7000 Datasheet HTML 19Page - Analog Devices ADPD7000 Datasheet HTML 20Page - Analog Devices ADPD7000 Datasheet HTML 21Page - Analog Devices ADPD7000 Datasheet HTML 22Page - Analog Devices ADPD7000 Datasheet HTML 23Page - Analog Devices Next Button
Zoom Inzoom in Zoom Outzoom out
 19 / 127 page
background image
Data Sheet
ADPD7000
THEORY OF OPERATION
analog.com
Rev. 0 | 19 of 127
the amount of headroom voltage at the LED driver pins, measured
with respect to ground, required to maintain the programmed LED
current levels. The compliance voltage is a function of the current
required.
ECG SIGNAL CHAIN
The ECG channel measures the differential voltage across two
electrodes to create a lead measurement. The output of this chan-
nel is a 24-bit digital word representing the measured ECG voltage.
The maximum input differential signal is ±1.2 V, and the LSB size is
286 nV. The output data rate can be 250 SPS, 500 SPS, 1 kSPS, 2
kSPS, or 4 kSPS. This filter selection affects the digital processing
but not the analog processing. Additionally, a 4 kHz AC lead off
signal is converted by the analog ECG path and extracted in the
digital domain.
The ECG channel has a dedicated ADC path with feedback ar-
rangement to remove the DC offset presented by the ECG electro-
des. The ECG channel is designed to provide a high quality ECG
signal process while suppressing the large DC offset that is caused
by the complex system design.
Both DC and AC lead off detection and DC lead on detection are in-
tegrated to accommodate different complex lead contact conditions
to provide reliable lead information.
The RLD signal is designed to better bias the human body potential
to avoid interference.
ECGIP and ECGIN are the signal inputs for the ECG channel, and
these inputs must be connected to the input leads. The RLD signal
must be connected to the reference lead.
The ECG signals (RLD, ECGIP, and ECGIN) can be routed to each
one of the ECG1 to ECG4 pins, controlled by settings within the
ECG_MATRIX register.
ECG Main Signal Chain
Figure 30 shows the ECG main signal chain diagram.
Figure 30. ECG Main Signal Chain Diagram
DC Lead Off (DCLO) Detection
The DCLO circuit injects a small programmable DC current into
each input electrode and monitors the resulting electrode voltage.
Figure 31 shows the DCLO detection diagram.
Figure 31. DCLO Detection Diagram
When both the ECGIP and ECGIN signals are properly connected,
the current flow from one of the ECG inputs through the patient
and into the other ECG input produces a minimal voltage shift. The
fail current magnitude is common across all ECG electrodes and
programmable from 0 nA to 1.1 μA.
The polarity of the fail current is independent for each elec-
trode and programmed using the ECG_DCLO_POLARITY_IN and
ECG_DCLO_POLARITY_IP bits (see Register 0x101). When an
electrode connection degrades or falls off, the signal voltage of that
electrode is pulled high or low depending on the programmed fail
current polarity.
The detection circuit is based on each individual electrode input
(ECGIP and ECGIN). The detection circuit uses comparators based
with independent programmable threshold levels for ECG inputs.
For each input lead (ECGIP and ECGIN), there is only one compa-
rator with a high-side threshold. For the RLD signal, there are two
comparators with a high-side threshold and low-side threshold.
These threshold limits are chosen such that all the threshold
voltages cover the expected signal range. The window compara-
tor compares the electrode input voltage with the corresponding
threshold voltages. When the voltage change on a particular elec-
trode exceeds one of the programmed threshold voltages for that
fail current, the DC resistance of the electrode contact can flag a
lead off.
AC Lead Off (ACLO) Detection
The ECG path has an ACLO detection circuit that can be used for
lead off detection with the DCLO detection circuit to deal with differ-
ent types of lead. Figure 32 shows the ACLO detection diagram.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100  ...More


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com