Back Button ADPD7000 Datasheet HTML 16Page - Analog Devices ADPD7000 Datasheet HTML 17Page - Analog Devices ADPD7000 Datasheet HTML 18Page - Analog Devices ADPD7000 Datasheet HTML 19Page - Analog Devices ADPD7000 Datasheet HTML 20Page - Analog Devices ADPD7000 Datasheet HTML 21Page - Analog Devices ADPD7000 Datasheet HTML 22Page - Analog Devices ADPD7000 Datasheet HTML 23Page - Analog Devices ADPD7000 Datasheet HTML 24Page - Analog Devices Next Button
Zoom Inzoom in Zoom Outzoom out
 20 / 127 page
background image
Data Sheet
ADPD7000
THEORY OF OPERATION
analog.com
Rev. 0 | 20 of 127
Figure 32. ACLO Detection Diagram
The ACLO method of sensing if the electrodes are connected to the
patient is based on injecting AC currents into each ECG electrode
and measuring the amplitudes of the resulting voltages through the
ECG channel. The ECG channel must be enabled to use the ACLO
function.
The magnitude of the ACLO current is programmable through the
ECG_ACLO_MAG bits. The AC current is driven out onto the ECG
input paths and establishes a voltage between the ECG channel in-
puts where it is measured. The measured voltage is synchronously
demodulated and sent through a comparator. The user can set the
threshold though the ECG_ACLO_THRESHOLD bits.
The ACLO detection functions only on the ECG input pins (ECGIP
and ECGIN) and is not supported for the RLD signal. A properly
connected electrode has a small signal because the drive current
flows into the right leg (RL). An improperly connected or dried out
electrode has a larger signal as determined by a capacitive voltage
divider.
Right Leg Driver (RLD)
The use of a driven reference benefits overall performance by
improving common-mode rejection of noise and interference from
external sources, such as power line interference (50 Hz or 60 Hz)
or other patient connected instruments. The drive stage also acts to
absorb lead fail currents injected into the ECG electrodes.
The reference electrode drive circuit senses the VCM of the patient
and drives an inverted version back to the body, creating a negative
feedback loop around the patient. The RLD amplifier uses VCM =
AVDD3/2, which centers the electrode voltages in the middle of the
ADC input range. Each electrode input is buffered and fed to the
RLD amplifier through a switch.
The amount of capacitance on the RLD signal affects the RLD
amplifier. For best performance, the capacitance on this node must
be less than 2.2 nF. In normal operation, the RLD amplifier output
is applied to the RL electrode via the RLD signal and associated
protection network.
The RLD amplifier can be turned off and the user can drive the RLD
signal by selecting a voltage in the ECG_RLD_OUT_SEL bits.
BIA SIGNAL CHAIN
The BIA signal chain is designed for body impedance measure-
ment. The signal chain consists of a high frequency precision
excitation loop and a measurement channel, which enables a wide
capability of measurement of the different bioimpedance configura-
tions.
The excitation loop consists of a 12-bit DAC, referred to as the
high speed DAC. This DAC is capable of generating high frequency
excitation signals up to 250 kHz.
The measurement channel features an ADC with input buffers, a
built-in antialias filter, and a PGA.
An ultralow leakage, programmable switch matrix connects the sen-
sor to the internal analog excitation and measurement blocks. This
matrix provides an interface for connecting external bioimpedance
and calibration resistors. Figure 33 shows the block diagram of this
BIA signal chain.
Figure 33. BIA Channel Block Diagram (ZCONT Is Contact Impedance)
The input multiplexer (mux) of the excitation loop is controlled
by the BIOZ_TSW_x, BIOZ_DSW_x, and BIOZ_RINT_SW_x bits.
The input mux for the measurement path is controlled by the
BIOZ_NCHAN_x and BIOZ_PCHAN_x. Table 10 shows the con-
nections of the input mux with the related bit setting. The flexible
design of the BIA channel allows different configurations of the
measurement.
Table 10. TIA Connections in the BIA Channel
Bit Name
Setting
Connection
BIOZ_TSW_x
0001
IMPIP
0010
IMPIN
0100
EXCP
1000
EXCN
BIOZ_RINT_SW_x, Bits[1:0]
1
The internal resistor (RINT)
that connects to the external
buffer (EXCBUF) and the
high power TIA (HPTIA)
blocks
0
RINT no connection
BIOZ_DSW_x
0001
IMPIP
0010
IMPIN
0100
EXCP
1000
EXCN
BIOZ_NCHAN_x
000
HPTIA_n1
001
IMPIN



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100  ...More


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com