Back Button MAX39 Datasheet HTML 7Page - Analog Devices MAX39 Datasheet HTML 8Page - Analog Devices MAX39 Datasheet HTML 9Page - Analog Devices MAX39 Datasheet HTML 10Page - Analog Devices MAX39 Datasheet HTML 11Page - Analog Devices MAX39 Datasheet HTML 12Page - Analog Devices MAX39 Datasheet HTML 13Page - Analog Devices MAX39 Datasheet HTML 14Page - Analog Devices MAX39 Datasheet HTML 15Page - Analog Devices Next Button
Zoom Inzoom in Zoom Outzoom out
 11 / 29 page
background image
www.analog.com
Analog Devices │ 11
MAX30009EVKIT#
Evaluation Kit
Evaluates: MAX30009
BioZ Mux Tab
The
BioZ Mux Tab (See Figure 9) configures the mux
that assigns pins to
DRVP, BIP, BIN, and DRVN to sup-
port various BioZ applications.
Wiring of MAX30009EVKIT# EV Kit Board
The MAX30009EVKIT# EV Kit board dedicates pins
EL2A
and
EL3A as circuit guards for pins EL2B and EL3B so
EL2A and EL3A cannot be selected by the mux for elec-
trode connections. In Tetrapolar electrode applications,
pins EL2B and EL3B should be connected to BIP and BIN
in the mux so that the
External Parasitic Cancellation
Enable in the BioZ Receive tab functions correctly.
Mux Enable
Enabling the mux connects electrode pins as selected
by
DRVP Assign, BIP Assign, BIN Assign, and DRVN
Assign. Disabling the mux disconnects all electrode and
calibration pins.
DRVN Out of Range Enable
The MAX30009 BioZ channel can perform DRVN lead-off
detection by monitoring the voltage of DRVN and trigger-
ing the BIOZ_DRV_OOR status bit when either the DRVP
or DRVN lead is disconnected. This is useful when using
a tetrapolar electrode configuration. Selecting
DRVN Out
of Range Enable enables DRVN lead-off detection. This
function can also be enabled in the
Advanced Settings:
Enable Drive Our of Range (OOR) Detection of the
BioZ Drive tab. DRVN Out of Range Enable is also
useful to detect if the Current magnitude selected in the
BioZ Drive tab is too large for the combined impedance of
the DRVP and DRVN electrode-tissue interfaces and the
body impedance.
Lead Bias
Lead-biasing drives the electrodes within the input
common-mode voltage requirements of the BioZ chan-
nel. It is recommended to enable both the BIP and BIN
Resistive Bias for tetrapolar measurements when any of
the Ex_C jumpers are open.
Figure 9. MAX30009EVKIT# EV Kit BioZ Mux Tab



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com