Back Button MAX39 Datasheet HTML 9Page - Analog Devices MAX39 Datasheet HTML 10Page - Analog Devices MAX39 Datasheet HTML 11Page - Analog Devices MAX39 Datasheet HTML 12Page - Analog Devices MAX39 Datasheet HTML 13Page - Analog Devices MAX39 Datasheet HTML 14Page - Analog Devices MAX39 Datasheet HTML 15Page - Analog Devices MAX39 Datasheet HTML 16Page - Analog Devices MAX39 Datasheet HTML 17Page - Analog Devices Next Button
Zoom Inzoom in Zoom Outzoom out
 13 / 29 page
background image
www.analog.com
Analog Devices │ 13
MAX30009EVKIT#
Evaluation Kit
Evaluates: MAX30009
The GSR test resistors are a selection of four high imped-
ance resistors meant for GSR/EDA applications.
For external calibration, a calibration resistor must be
connected either to the CALx pins or to the ELx electrode
outputs. The resistor value must be entered into the
External Resistor Value box and the Set button must be
pressed to update the value.
To use the CALx pins, select
External Resistor on CALX
pins, which connects CAL1 pin to DRVP, CAL2 pin to
BIP, CAL3 pin to BIN, and CAL4 pin to DRVN through
the BioZ Mux. The
CALx pins are brought out to the
CALIBRATION PORT (connector J10) which can be con-
nected to the calibration resistor. Pins CAL1 and CAL2
should be connected to one side of the resistor with CAL3
and CAL4 connected to the other side of the resistor.
The
Connect ELx Pins during CALx Calibration check-
box provides the option of connecting the ELx pins in
addition to the CALx pins through the BioZ Mux, which is
designed to add the ELx parasitic effects during the CALx
calibration.
An external resistor can also be connected between the
pins of the electrodes. Select External Resistor
on BIOZ_
ELx Pins and connect the resistor between the DRVP/
BIP and DRVN/BIN pins.
The
One-Shot Measure button can be used to get a
quick measurement of voltage and impedance using the
settings defined in BioZ Drive, BioZ Receive, and BioZ
Mux without having to start data acquisition and having
to view measurements in the
Plots tab. This is useful for
calibration.
The
Frequency Calibration box contains the functions to
set up the calibration frequencies.
●
Import: Loads frequency calibration CSV file previ-
ously saved by
Export function.
●
Export: Saves the frequency calibration setup to a
CSV file for future use.
●
Add: Reads the settings from the PLL tab and adds
an entry to the calibration frequency selection table
with a specific drive frequency and sample rate.
●
Load: After the user selects a specific frequency cali-
bration setting line,
Load programs the PLL registers
based on the Drive Freq and ADC SR values in the
line.
●
Remove: Removes a line in the calibration frequency
selection table. To select a line, click on any field of
that specific line in the calibration frequency selection
table.
●
Remove All: Clears the calibration frequency selec-
tion table.
●
Settling Points: Number of samples that are taken
but not used, to allow the impedance measurement
to settle.
●
Average Points: Number of impedance samples
averaged to derive calibration values.
●
Calibrate: Performs the calibration at each frequency,
calculates the coefficients, and offsets and populates
those fields with the calculated values.
●
Enable Software Sweep Mode Operation (Q vs. I
Plot): This mode performs an automated measure-
ment sweep at the frequencies defined in the table.
The measurement timing is defined by the
Settling
Points and Average Points. Calibration is automati-
cally applied to the measurements, and the calibrated
data is displayed on the
Bioz (Q vs. I) plot on the
Plots tab.
●
Apply Calibration to BioZ vs. Time Plot: Selecting
this option applies the calibration at the loaded fre-
quency to the
BioZ, BioZ (Mag/Ph), and Combination
plots in the
Plots tab. This is useful to observe
calibrated time-domain signals.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com