12 / 29 page ![]() www.analog.com Analog Devices │ 12 MAX30009EVKIT# Evaluation Kit Evaluates: MAX30009 DC Lead-Off Detect The MAX30009 BioZ channel can perform DC lead-off detection by injecting DC current and comparing the dif- ferential voltage across BIP and BIN to the voltage thresh- olds selected in the Threshold. The DC source and sink currents are defined in the settings Current Magnitude and Current Polarity. The External DC Lead Off func- tion is not supported by the MAX30009EVKIT# EV kit board wiring. Refer to the MAX30009 datasheet under the Leads Off Detection section for more details. BioZ Calibration Tab The BioZ Calibration Tab is shown in Figure 10. The MAX30009 BioZ channel can be calibrated to ensure accurate impedance measurements. For proper calibra- tion, gain and phase errors need to be calculated for each stimulus frequency. Refer to the MAX30009 IC datasheet under the BioZ Calibration section for a full description of the calibration procedure and equations to calculate gain and phase errors. The calibration procedure uses either internal programma- ble resistors or an external calibration resistor connected at the calibration pins or at the electrode pins. The resistor options are selected in the Calibration Resistor box. Only one option (external or internal) should be selected. For internal calibration, the MAX30009 provides two inter- nal programmable resistors: BioZ test resistors and GSR test resistors. When these internal resistors are used the Mux Enable in the BioZ Mux tab should be deselected to avoid external interference, and only one internal resistor type (BioZ or GSR) should be selected. The BioZ test resistors are a selection of low impedance resistors meant for BIA/BIS applications. These resis- tors have nominal values, but due to tolerance errors a measurement is performed during the device test and the resistor error correction value is stored as the BIST_R_ ERR in register 0x44. The GUI uses this value to calculate and display the Actual Resistance (Error Corrected) value for use in calibration calculations. Figure 10. MAX30009EVKIT# EV Kit BioZ Calibration Tab |
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |