Back Button AD5683R Datasheet HTML 2Page - Analog Devices AD5683R Datasheet HTML 3Page - Analog Devices AD5683R Datasheet HTML 4Page - Analog Devices AD5683R Datasheet HTML 5Page - Analog Devices AD5683R Datasheet HTML 6Page - Analog Devices AD5683R Datasheet HTML 7Page - Analog Devices AD5683R Datasheet HTML 8Page - Analog Devices AD5683R Datasheet HTML 9Page - Analog Devices AD5683R Datasheet HTML 10Page - Analog Devices Next Button
Zoom Inzoom in Zoom Outzoom out
 6 / 12 page
background image
CN-0411
Circuit Note
Rev. 0 | Page 6 of 12
Temperature Compensation
Conductivity measurements are also temperature dependent;
conductance increases as temperature increases. Most
commercial conductivity probes have an integrated RTD,
facilitating temperature compensation.
The typical method to compensate the conductivity
measurement to a standard reference temperature is through a
linear function dependent on the temperature coefficient α,
which is based on the type of ions present in the solution as
shown in Equation 10. Usually, the reference temperature is set
at 25 ˚C.
(
)
1
REF
REF
Y
Y
TT
=
+α
−
(10)
where:
YREF is the conductivity referenced to TREF.
Y is the conductivity at temperature T.
TREF is the reference temperature.
T is the solution temperature.
α is the temperature coefficient.
The typical values of α for common salt solutions are shown in
Table 2.
Table 2. Temperature Coefficients for Common Salt
Solutions
Salt Solution
Temperature Coefficient (α)
Potassium Chloride (KCl)
1.88
Sodium Chloride (NaCl)
2.14
The TDS of the solution in mg/L is calculated from the referenced
conductivity measurement using Equation 11.
TDS = ke × YREF
(11)
where ke is the TDS factor.
The TDS factor is calculated by measuring the referenced
conductivity of a standard known TDS solution as shown in
Equation 12.
,
,
REF STD
e
REF STD
TDS
k
Y
=
(12)
where:
TDSREF,STD is the known TDS value of the solution at TREF.
YREF,STD is the measured conductivity of the solution at TREF.
The conductivity measurement cannot distinguish between
constituents and individual types of ions. Thus, the TDS factor
varies at a specific range, which is distinct for each type of
solution. Table 3 shows the typical range of the TDS factors of
common salt solutions.
Table 3. TDS Factors of Common Salt Solutions
Salt Solution
Range of TDS Factor (ke)
Potassium Chloride (KCl)
0.50 to 0.57
Sodium Chloride (NaCl)
0.47 to 0.50
Temperature Measurement
Conductivity varies significantly with the temperature of the
solution and the temperature coefficients are also significantly
for each type of solution. A simple method of measuring the
temperature of a solution uses a 2-wire RTD. The simplified
front-end schematic is shown in Figure 10.
Figure 10. 2-Wire RTD Simplified Schematic
The AD7124-8 contains two matched software configurable
constant current sources, which have eight selectable current
output values and can be made available to any of the analog
input channels. The excitation current to the RTD resistor
network is set to 250 µA from the AIN0 channel. Node A and
Node B are connected to the AD7124-8 external reference
input. Node B and Node C, which constitute the differential
RTD signal, are connected to the AD7124-8 analog input
channels. Both differential inputs are passed through two
identical and typical RC low-pass filter as shown in Figure 11.
Figure 11. Typical RC Low-Pass Filter for Differential Inputs
This low-pass filter topology provides attenuation of both
differential and common-mode noise signals. The large value of
R1 and R2 in Figure 11 also help to protect against 30 V
miswiring.C1 and C2 are the common-mode capacitors and are
set to a tenth of the differential-mode capacitor C3. This
attenuates the effects of noise due to mismatch between the
common-mode capacitors. The common mode cut-off
frequency (fCM) is approximately 20 times the differential mode
cut-off frequency (fDM), which are 53 kHz and 2.5 kHz,
respectively. The calculations are shown in Equation 13.
(
)
1
1
; for 
 and 
22
1
; for 
 or 
2
2
CM
1
2
1
2
11
2 2
DM
1
2
12
12
3
f
R = R
C =C
RC
R C
f
C =C
CC
R +R
C
=
=
ππ
=
π+
(13)
where:
fCM is the common mode cut-off frequency.
fDM is the differential mode cut-off frequency.
All resistance values for the RTD are referenced or capped to
4.02 kΩ and the reference voltage is 250 µA × 4.02 kΩ = 1.005 V.
A
4.02kΩ
C
RTD
IEXC
1.5kΩ
B
POSITIVE
ADC INPUT
NEGATIVE
ADC INPUT
POSITIVE
SIGNAL INPUT
NEGATIVE
SIGNAL INPUT
R1
3kΩ
R2
3kΩ
C1
1nF
C2
1nF
C3
10nF



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com