Back Button ADPD4000 Datasheet HTML 21Page - Analog Devices ADPD4000 Datasheet HTML 22Page - Analog Devices ADPD4000 Datasheet HTML 23Page - Analog Devices ADPD4000 Datasheet HTML 24Page - Analog Devices ADPD4000 Datasheet HTML 25Page - Analog Devices ADPD4000 Datasheet HTML 26Page - Analog Devices ADPD4000 Datasheet HTML 27Page - Analog Devices ADPD4000 Datasheet HTML 28Page - Analog Devices ADPD4000 Datasheet HTML 29Page - Analog Devices Next Button
Zoom Inzoom in Zoom Outzoom out
 25 / 82 page
background image
Data Sheet
ADPD4000/ADPD4001
Rev. A | Page 25 of 82
APPLICATIONS INFORMATION
OPERATING MODE OVERVIEW
The ADPD4000/ADPD4001 are effectively charge measuring
devices that can interface with many different sensors enabling
synchronous measurements of PPG, electrocardiography
(ECG), electrodermal activity (EDA), impedance, capacitance,
and temperature measurements. A selection of operating modes
are built into the device to optimize each of the different sensor
measurements supported.
SINGLE INTEGRATION MODE
Single integration mode is used for a single integration of
incoming charge per ADC conversion and is the most common
operating mode for the ADPD4000/ADPD4001. In single
integration mode, most of the dynamic range of the integrator is
used when integrating the charge from the sensor response to a
single stimuli event, for example, an LED pulse. There is also a
multiple integration mode available for situations with very
small sensor responses (see the Multiple Integration Mode
section for more information).
Using LED as Stimulus
Single integration mode is the typical operating mode used for a
PPG measurement, where an LED is pulsed into human tissue
and the resultant charge from the photodiode response is
integrated and subsequently converted by the ADC. Figure 32
shows an example of a typical PPG measurement circuit.
TIA
ADC
BPF
RF
RINT
6.3pF
6.3pF
TIA_VREF
SWITCH
VLED1
CVLED
VCx
INx
LEDx
RF
RINT
Figure 32. Typical PPG Measurement Circuit
The MOD_TYPE_x value is left at the default value of 0 so that
the TIA is continuously connected to the input of the TIA. Set
the PRECON_x bit field to 0x5 to set the anode of the photodiode
(PD) to the TIA_VREF potential during the preconditioning
period. The VCx pin is connected to the cathode of the
photodiode and is set to TIA_VREF + 250 mV to apply a
250 mV reverse bias across the photodiode, which reduces the
photodiode capacitance and reduces the noise of the signal
path. Set TIA_VREF to 1.27 V using the AFE_TRIM_VREF_x
bit field for maximum dynamic range.
The LED pulse is controlled with the LED_OFFSET_x and LED_
WIDTH_x bit field. The default LED offset (LED_OFFSET_x =
0x10) is 16 μs from the end of the preconditioning period and is
suitable for most use cases. Recommended LED pulse widths
are either 2 μs or 3 μs when using the BPF. Shorter LED pulse
widths provide the greatest amount of ambient light rejection
and the lowest power dissipation. The period is automatically
calculated by the ADPD4000/ADPD4001. The automatic
calculation is based on the integration width selected and the
number of ADC conversions. To use the automatic calculation,
leave the MIN_PERIOD_x bit field at its default value of 0. If a
longer period is desired, for example, if more settling time is
required, use the MIN_PERIOD_x bit field to enable a longer
period.
The integration pulses are controlled with the INTEG_
OFFSET_x, INTEG_FINE_OFFSET_x, and INTEG_WIDTH_x
bit fields. It is recommended that an integration width of 1 μs
greater than the LED width be used because the signal spreads
due to the response of the BPF. By setting the integration width
1 μs wider than the LED width, a maximum amount of charge
from the incoming signal is integrated.
The number of ADC conversions defaults to a single ADC
conversion. However, oversampling is available for increased
SNR. The ADC conversions can be set to 1, 2, 3, or 4, based on
the ADC_COUNT_x bit field. If two channels are enabled,
Channel 1 occurs first, followed by Channel 2. The total number of
pulses is equal to NUM_INT_x × NUM_REPEAT_x. In single
integration mode, NUM_INT_x = 1 for a single integration
sequence per ADC conversion. Therefore, the total number of
pulses is controlled by NUM_REPEAT_x. Increasing the
number of pulses reduces the noise floor of the measurement by a
factor of √n, where n is the total number of pulses.
Figure 33 shows the timing operation where a single integration
cycle is used per ADC conversion. Table 17 details the relevant
registers using single integration mode for a PPG measurement.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com