Back Button PMP75 Datasheet HTML 13Page - Endress+Hauser Group Services AG PMP75 Datasheet HTML 14Page - Endress+Hauser Group Services AG PMP75 Datasheet HTML 15Page - Endress+Hauser Group Services AG PMP75 Datasheet HTML 16Page - Endress+Hauser Group Services AG PMP75 Datasheet HTML 17Page - Endress+Hauser Group Services AG PMP75 Datasheet HTML 18Page - Endress+Hauser Group Services AG PMP75 Datasheet HTML 19Page - Endress+Hauser Group Services AG PMP75 Datasheet HTML 20Page - Endress+Hauser Group Services AG PMP75 Datasheet HTML 21Page - Endress+Hauser Group Services AG Next Button
Zoom Inzoom in Zoom Outzoom out
 17 / 124 page
background image
Cerabar S PMC71, PMP71, PMP75
Endress+Hauser
17
I
63 %
100 %
t
t
1
t
2
90 %
t
3
A0019786
Dynamic behavior,
current output
 
Type
Measuring cell
Dead time (t1) [ms]
Time constant T63 (t2) [ms]
Time constant T90 (t3) [ms]
PMC71
max.
All
90
120
276
PMP71
max.
• 400 mbar (6 psi)
• ≥1 bar (15 psi)
45
• 70
• 35
• 161
• 81
PMP75
max.
PMP71 + influence of the diaphragm seal
Dynamic behavior
digital output (
HART electronics)
A typical burst rate of 300 ms results in the following behavior:
 
Type
Measuring cell
Dead time (t1) [ms]
Dead time (t1) [ms] +
Time constant T63 (t2) [ms]
Dead time (t1) [ms] +
Time constant T90 (t3) [ms]
PMC71
Min.
All
250
370
436
max.
1050
1170
1236
PMP71
Min.
• 400 mbar (6 psi)
• ≥1 bar (15 psi)
205
• 275
• 240
• 321
• 241
max.
• 400 mbar (6 psi)
• ≥1 bar (15 psi)
1005
• 1075
• 1040
• 1121
• 1041
PMP75
max.
PMP71 + influence of the diaphragm seal
Reading cycle
• Acyclic: max. 3/s, typically 1/s (depending on command # and number of preambles)
• Cyclic (burst): max. 3/s, typically 2/s
The device commands the BURST MODE functionality for cyclical value transmission via the HART
communication protocol.
Cycle time (update time)
Cyclic (burst): min. 300 ms
Response time
• Acyclic: min. 330 ms, typically 590 ms (depending on command # and number of preambles)
• Cyclic (burst): min. 160 ms, typically 350 ms (depending on command # and number of
preambles)
Dynamic behavior
PROFIBUS PA
A typical PLC cycle time of 1 s results in the following behavior:
Type
Measuring cell
Dead time (t1) [ms]
Dead time (t1) [ms] +
Time constant T63 (t2) [ms]
Dead time (t1) [ms] +
Time constant T90 (t3) [ms]
PMC71
Min.
All
125
245
311
max.
1325
1445
1511



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100  ...More


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com