Back Button ADS1291IRSMT Datasheet HTML 43Page - Texas Instruments ADS1291IRSMT Datasheet HTML 44Page - Texas Instruments ADS1291IRSMT Datasheet HTML 45Page - Texas Instruments ADS1291IRSMT Datasheet HTML 46Page - Texas Instruments ADS1291IRSMT Datasheet HTML 47Page - Texas Instruments ADS1291IRSMT Datasheet HTML 48Page - Texas Instruments ADS1291IRSMT Datasheet HTML 49Page - Texas Instruments ADS1291IRSMT Datasheet HTML 50Page - Texas Instruments ADS1291IRSMT Datasheet HTML 51Page - Texas Instruments Next Button
Zoom Inzoom in Zoom Outzoom out
 47 / 83 page
background image
47
ADS1291, ADS1292, ADS1292R
www.ti.com
SBAS502C – DECEMBER 2011 – REVISED APRIL 2020
Product Folder Links: ADS1291 ADS1292 ADS1292R
Submit Documentation Feedback
Copyright © 2011–2020, Texas Instruments Incorporated
(1)
When in RDATAC mode, the RREG command is ignored.
(2)
n nnnn = number of registers to be read or written – 1. For example, to read or write three registers, set n nnnn = 0 (0010). r rrrr =
starting register address for read and write opcodes.
8.5.2 SPI Command Definitions
The ADS1291, ADS1292, and ADS1292R provide flexible configuration control. The opcode commands
summarized in Table 15 control and configure the ADS1291, ADS1292, and ADS1292R operation. The opcode
commands are stand-alone, except for the register read and register write operations that require a second
command byte plus data. CS can be taken high or held low between opcode commands but must stay low for
the entire command operation (especially for multi-byte commands). System opcode commands and the RDATA
command are decoded by the ADS1291, ADS1292, and ADS1292R on the seventh SCLK falling edge. The
register read and write opcodes are decoded on the eighth SCLK falling edge. Be sure to follow SPI timing
requirements when pulling CS high after issuing a command.
Table 15. Command Definitions
COMMAND
DESCRIPTION
FIRST BYTE
SECOND BYTE
System Commands
WAKEUP
Wake-up from standby mode
0000 0010 (02h)
STANDBY
Enter standby mode
0000 0100 (04h)
RESET
Reset the device
0000 0110 (06h)
START
Start or restart (synchronize) conversions
0000 1000 (08h)
STOP
Stop conversion
0000 1010 (0Ah)
OFFSETCAL
Channel offset calibration
0001 1010 (1Ah)
Data Read Commands
RDATAC
Enable Read Data Continuous mode.
This mode is the default mode at power-up.(1)
0001 0000 (10h)
SDATAC
Stop Read Data Continuously mode
0001 0001 (11h)
RDATA
Read data by command; supports multiple read back.
0001 0010 (12h)
Register Read Commands
RREG
Read n nnnn registers starting at address r rrrr
001r rrrr (2xh)(2)
000n nnnn(2)
WREG
Write n nnnn registers starting at address r rrrr
010r rrrr (4xh)(2)
000n nnnn(2)
8.5.2.1 WAKEUP: Exit STANDBY Mode
This opcode exits the low-power standby mode; see the STANDBY: Enter STANDBY Mode subsection of the
SPI Command Definitions section. Time is required when exiting standby mode (see the Electrical
Characteristics for details). There are no restrictions on the SCLK rate for this command and it can be
issued any time. Any following command must be sent after 4 tCLK cycles.
8.5.2.2 STANDBY: Enter STANDBY Mode
This opcode command enters the low-power standby mode. All parts of the circuit are shut down except for the
reference section. The standby mode power consumption is specified in the Electrical Characteristics. There are
no restrictions on the SCLK rate for this command and it can be issued any time. Do not send any other
command other than the wakeup command after the device enters the standby mode.
8.5.2.3 RESET: Reset Registers to Default Values
This command resets the digital filter cycle and returns all register settings to the default values. See the Reset
(RESET) subsection of the SPI Interface section for more details. There are no restrictions on the SCLK rate
for this command and it can be issued any time. It takes 9 fMOD cycles to execute the RESET command.
Avoid sending any commands during this time.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com