Back Button AD4115 Datasheet HTML 27Page - Analog Devices AD4115 Datasheet HTML 28Page - Analog Devices AD4115 Datasheet HTML 29Page - Analog Devices AD4115 Datasheet HTML 30Page - Analog Devices AD4115 Datasheet HTML 31Page - Analog Devices AD4115 Datasheet HTML 32Page - Analog Devices AD4115 Datasheet HTML 33Page - Analog Devices AD4115 Datasheet HTML 34Page - Analog Devices AD4115 Datasheet HTML 35Page - Analog Devices Next Button
Zoom Inzoom in Zoom Outzoom out
 31 / 51 page
background image
Data Sheet
AD4115
Rev. 0 | Page 31 of 51
STANDBY MODE AND POWER-DOWN MODE
In standby mode, most blocks are powered down. The LDO
regulators remain active so that the registers maintain the
individual register contents. The crystal oscillator remains
active if selected. To power down the clock in standby mode,
set the CLOCKSEL bits in the ADC mode register to 00
(internal oscillator).
In power-down mode, all blocks are powered down, including
the LDO regulators. All registers lose the individual register
contents and the GPIO outputs are placed in three-state. To
prevent accidental entry into power-down mode, place the
ADC in standby mode first. Exiting power-down mode
requires a serial interface reset (64 serial clocks with CS = 0 and
DIN = 1). A delay of 500 µs is recommended before issuing a
subsequent serial interface command to allow the LDO
regulator to power up.
CALIBRATION MODES
The AD4115 allows the user to perform a two-point calibration
to eliminate any offset and gain errors. The following four
calibration modes are used to eliminate these offset and gain
errors on a per setup basis:
•
Internal zero-scale calibration mode
•
Internal full-scale calibration mode
•
System zero-scale calibration mode
•
System full-scale calibration mode
Only one channel can be active during calibration. After each
conversion, the ADC conversion result is scaled using the ADC
calibration registers before being written to the data register.
The default value of the offset register is 0x800000, and the
default value of the gain register is 0x5XXXX0. The following
equations show the calculations used to scale the ADC conversion
result. In unipolar mode, the ideal relationship (not considering
the ADC gain error and offset error) is calculated as follows:
Data = ((0.075 × VIN/VREF) × 223 – (Offset − 0x800000)) ×
(Gain/0x400000) × 2
In bipolar mode, the ideal relationship (not considering the
ADC gain error and offset error) is calculated as follows:
Data = ((0.075 × VIN/VREF) × 223 – (Offset − 0x800000)) ×
(Gain/0x400000) + 0x800000
To start a calibration, write the relevant value to the Mode bits
in the ADC mode register Bits[6:4]. The DOUT/RDY pin and
the RDY bit in the status register go high when the calibration
initiates. When the calibration is complete, the contents of the
corresponding offset or gain register update, the RDY bit in the
status register resets, the RDY output pin returns low (if CS is
low), and the AD4115 reverts to standby mode.
During an internal offset calibration both modulator inputs are
connected internally to the selected negative analog input pin,
and the user must ensure that the voltage on the selected negative
analog input pin does not exceed the allowed limits and is free
from excessive noise and interference. A full-scale input voltage
automatically connects to the ADC input to perform an
internal full-scale calibration.
For system calibrations, apply the system zero-scale (offset) and
system full-scale (gain) voltages to the input pins before
initiating the calibration modes. As a result, errors external to
the AD4115 are removed. The calibration range of the ADC
gain for a system full-scale calibration on a voltage input is
from 3.75 × VREF to 10.5 × VREF. If 10.5 × VREF is greater than the
absolute input voltage specification for the applied AVDD, use the
specification as the upper limit instead of 10.5 × VREF (see the
Specifications section).
An internal zero-scale calibration only removes the offset error
of the ADC core. This calibration does not remove error from
the resistive front end. A system zero-scale calibration reduces
the offset error to the order of the noise on that channel.
From an operational point of view, treat a calibration like
another ADC conversion. Always perform an offset calibration,
if required, before a full-scale calibration. Set the system
software to monitor the RDY bit in the status register or the
RDY output to determine the end of a calibration via a polling
sequence or an interrupt driven routine. All calibrations require
a time equal to the settling time of the selected filter and output
data rate to be completed.
Any calibration can be performed at any output data rate.
Lower output data rates result in higher calibration accuracy
and the calibration is accurate for all output data rates. A new
offset calibration is required for a given channel if the reference
source for that channel is changed.
The AD4115 provides access to the on-chip offset and gain
calibration registers to allow the microprocessor to read the
device calibration coefficients and to write the stored
calibration coefficients. A read or write of the offset and gain
registers can be performed at any time except during an
internal or self calibration.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com