Back Button ADS1294CZXGR Datasheet HTML 25Page - Texas Instruments ADS1294CZXGR Datasheet HTML 26Page - Texas Instruments ADS1294CZXGR Datasheet HTML 27Page - Texas Instruments ADS1294CZXGR Datasheet HTML 28Page - Texas Instruments ADS1294CZXGR Datasheet HTML 29Page - Texas Instruments ADS1294CZXGR Datasheet HTML 30Page - Texas Instruments ADS1294CZXGR Datasheet HTML 31Page - Texas Instruments ADS1294CZXGR Datasheet HTML 32Page - Texas Instruments ADS1294CZXGR Datasheet HTML 33Page - Texas Instruments Next Button
Zoom Inzoom in Zoom Outzoom out
 29 / 114 page
background image
ADS1294, ADS1294R, ADS1296, ADS1296R, ADS1298, ADS1298R
www.ti.com
SBAS459K – JANUARY 2010 – REVISED AUGUST 2015
Feature Description (continued)
9.3.1.3.5
Supply Measurements (MVDDP, MVDDN)
Setting CHnSET[2:0] = 011 sets the channel inputs to different supply voltages of the device.
For channels 1, 2, 5, 6, 7, and 8, (MVDDP – MVDDN) = [0.5 × (AVDD – AVSS)]
For channels 3 and 4, (MVDDP – MVDDN) = DVDD / 4.
To avoid saturating the PGA while measuring power supplies, set the gain to 1.
For example, if AVDD = 2.5 V and AVSS = –2.5 V, then the measurement result is 2.5 V.
9.3.1.3.6
Lead-Off Excitation Signals (LoffP, LoffN)
The lead-off excitation signals are fed into the multiplexer before the switches. The comparators that detect the
lead-off condition are also connected to the multiplexer block before the switches. For a detailed description of
the lead-off block, refer to the Lead-Off Detection section.
9.3.1.3.7
Auxiliary Single-Ended Input
The RLD_IN pin is primarily used for routing the right leg drive (RLD) signal to any of the electrodes in case the
RLD electrode falls off. However, the RLD_IN pin can be used as a multiple single-ended input channel. The
signal at the RLD_IN pin can be measured with respect to the voltage at the RLD_REF pin using any of the eight
channels. This measurement is done by setting the channel multiplexer setting to 010, and the RLD_MEAS bit of
the CONFIG3 register to 1.
Copyright © 2010–2015, Texas Instruments Incorporated
Submit Documentation Feedback
29
Product Folder Links: ADS1294 ADS1294R ADS1296 ADS1296R ADS1298 ADS1298R



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100  ...More


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com