PMP71 Datasheet(HTML) 40 Page - Endress+Hauser Group Services AG |
|
40 / 124 page ![]() Cerabar S PMC71, PMP71, PMP75 40 Endress+Hauser Step 3: Calculation of the ambient temperature effect [E2 = (E2M · CF1) + E2E] → 31 For the measuring conditions, Main temperature error, E2M = ± (0.03 · TD + 0.03) % of the set span Temperature correction factor, CF1 = 1 Electronics error , E2E = 0.05 % Ambient temperature effect, E2 = ± [(0.03 · TD + 0.03) · 1] +0.05 = ± [(0.03 · 2 + 0.03) · 1] + 0.05 = ± 0.14 (% of the set span) = ± (0.14/100) · 5 bar (75 psi) = ±0.007 bar (0.1015 psi) E2 = ± 0.14 (% of the set span) (or) ±0.007 bar (0.1015 psi) A B C 0.0 0.05 0.1 0.15 0.2 0.25 0.3 [%] -40 -40 -30 -22 -10 +14 0 +32 +10 +50 +20 +68 +30 +86 +40 +104 +50 +122 +60 +140 +70 +158 +80 +176 +90 +194 -20 -4 [°C] [°F] D A0031299 A Error (% of the set span) B Temperature C 316L process isolating diaphragm D Ambient temperature effect: 0.14 (% of the set span) |
|
Link URL |
| Does ALLDATASHEET help your business so far? [ DONATE ] |
About Alldatasheet | Advertisement | Contact us | Privacy Policy | Link to Datasheet | Link Exchange | Manufacturer List All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|