Back Button PMP71 Datasheet HTML 104Page - Endress+Hauser Group Services AG PMP71 Datasheet HTML 105Page - Endress+Hauser Group Services AG PMP71 Datasheet HTML 106Page - Endress+Hauser Group Services AG PMP71 Datasheet HTML 107Page - Endress+Hauser Group Services AG PMP71 Datasheet HTML 108Page - Endress+Hauser Group Services AG PMP71 Datasheet HTML 109Page - Endress+Hauser Group Services AG PMP71 Datasheet HTML 110Page - Endress+Hauser Group Services AG PMP71 Datasheet HTML 111Page - Endress+Hauser Group Services AG PMP71 Datasheet HTML 112Page - Endress+Hauser Group Services AG Next Button
Zoom Inzoom in Zoom Outzoom out
 108 / 124 page
background image
Cerabar S PMC71, PMP71, PMP75
108
Endress+Hauser
Design and operation mode
Diaphragm seals are separating equipment between the measuring system and the process.
A diaphragm seal system consists of:
• A diaphragm seal
• If necessary, a capillary tube or temperature isolator
• Fill fluid
• A pressure transmitter
The process pressure acts via the process isolating diaphragm of a diaphragm seal on the liquid-filled
system, which transfers the process pressure onto the sensor of the pressure transmitter.
Endress+Hauser delivers all diaphragm seal systems as welded versions. The system is hermetically
sealed, which ensures the highest reliability.
The diaphragm seal determines the application range of the system by:
• The process isolating diaphragm diameter
• The process isolating diaphragm stiffness and material
• The design (oil volume)
diameter of the process isolating diaphragm
The greater the diameter of the process isolating diaphragm (less stiff), the smaller the temperature
effect on the measurement result.
Stiffness of the process isolating diaphragm
The stiffness depends on the diameter of the process isolating diaphragm, the material, any existing
coating and the thickness and shape of the process isolating diaphragm. The process isolating
diaphragm thickness and the shape are determined by the design. The stiffness of a process isolating
diaphragm of a diaphragm seal influences the temperature operating range and the measuring error
caused by temperature effects.
The new TempC membrane: Highest accuracy and process safety in pressure and differential pressure
measurement with diaphragm seals
To measure even more precisely in such applications and to increase process safety, Endress+Hauser
has developed the TempC membrane based on a completely revolutionary technology. This
diaphragm guarantees the highest level of accuracy and process safety in diaphragm seal
applications.
• The very low temperature effect minimizes the influence of fluctuations of both process and
ambient temperatures, thus guaranteeing accurate and stable measurements. Measurement
inaccuracies caused by temperature are reduced to a minimum.
• The TempC membrane can be used at temperatures between –40 °C (–40 °F) and
+250 °C (+482 °F). This guarantees maximum process safety even for very long sterilization and
cleaning cycles (SIP/CIP) in tanks and pipes at high temperatures.
• Thanks to the TempC membrane, smaller dimension process connections are possible. With a
smaller process connection, the new diaphragm measures at least as accurately as a conventional
diaphragm with a larger diameter.
• Short recovery times after temperature shocks mean less downtime during batch processes and
therefore a far higher level of availability of the production facilities.
• In addition, the TempC membrane excels in terms of its improved hygienic cleanability and its
insensitivity to substantial changes in the pressure load.
Ordering information:
See the Product Configurator for the individual process connection and the choice of process
isolating diaphragm.
Selection in the Applicator:
Under "Transmitter data" in the "Diaphragm material" field.
Capillary
As standard, capillaries with an internal diameter of 1 mm (0.04 in) are used.
The capillary tube influences the thermal change, the ambient temperature operating range and the
response time of a diaphragm seal system as a result of its length and internal diameter.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100  ...More


Datasheet Download

Go To PDF Page


Link URL



Does ALLDATASHEET help your business so far?  [ DONATE ] 

About Alldatasheet   |   Advertisement   |   Contact us   |   Privacy Policy   |   Link to Datasheet    |   Link Exchange   |   Manufacturer List
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com